Add like
Add dislike
Add to saved papers

Evaluation of Accuracy and Safety of the Next-Generation Up to 180-Day Long-Term Implantable Eversense Continuous Glucose Monitoring System: The PROMISE Study.

Background: Use of continuous glucose monitoring (CGM) systems is being rapidly adopted as standard of care for insulin-requiring patients with diabetes. The PROMISE study (NCT03808376) evaluated the accuracy and safety of the next-generation implantable Eversense CGM system for up to 180 days. Methods: This was a prospective multicenter study involving 181 subjects with diabetes at 8 USA sites. All subjects were inserted with a primary sensor. Ninety-six subjects had a second sensor, either an identical sensor or a modified sensor (sacrificial boronic acid [SBA]), inserted in their other arm (53 and 43 subjects, respectively). Accuracy was evaluated by comparing CGM to YSI 2300 glucose analyzer (Yellow Springs Instrument [YSI]) values during 10 clinic visits (day 1-180). Confirmed event detection rates, calibration stability, sensor survival, and serious adverse events (SAEs) were evaluated. Results: For primary sensors, the percent CGM readings within 20%/20% of YSI values was 92.9%; overall mean absolute relative difference (MARD) was 9.1%. The confirmed alert detection rate at 70 mg/dL was 93% and at 180 mg/dL was 99%. The median percentage of time for one calibration per day was 56%. Sixty-five percent of the primary sensors survived to 180 days. For the SBA sensors, the percent CGM readings within 20%/20% of YSI values was 93.9%; overall MARD was 8.5%. The confirmed alert detection rate at 70 mg/dL was 94% and at 180 mg/dL was 99%. The median percentage of time for one calibration per day was 63%. Ninety percent of the SBA sensors survived to 180 days. No device- or insertion/removal procedure-related SAEs were reported. Conclusion: These data show the next-generation Eversense CGM system had sustained accuracy and safety up to 180 days, with an improved calibration scheme and survival, using the primary or SBA sensors.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app