Add like
Add dislike
Add to saved papers

An unprecedented cause of cardiac resynchronization with defibrillator (CRT-D) malfunction "A beheaded generator assembly".

Cardiac resynchronization with defibrillator (CRT-D) malfunction can be due to various reasons, including wire injury, insulation break, battery problems, or patient factors. Undesired outcomes can be dramatically elevated in those settings, prompting early detection and proper troubleshooting. To diagnose, clinical correlation and device interrogation are vital. However, it is not uncommon to find general troubleshooting options insufficient, as highlighted in this case report. Here, we presented an unusual "head assembly separation," as the main reason for abnormal device parameters.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app