We have located links that may give you full text access.
Evaluating the incidence of spinal cord injury after spinal cord stimulator implant: an updated retrospective review.
Regional Anesthesia and Pain Medicine 2022 July
INTRODUCTION: While spinal cord stimulator implant is an increasingly safe procedure, one of the most feared complications is spinal cord injury. Still, literature regarding its incidence remains highly variable. This retrospective analysis aims to evaluate the incidence of spinal cord injury after spinal cord stimulator implant using a large-scale claims database.
METHODS: The PearlDiver-Mariner database of national all payer claims was used to identify patients who underwent spinal cord stimulator implant (percutaneous or paddle) and developed subsequent spinal cord injury within 45 days. The primary outcome was to determine the overall incidence of spinal cord injury after spinal cord stimulator implant. Secondary outcomes included an evaluation of potential factors associated with developing spinal cord injury using univariable and multivariable regression analysis.
RESULTS: A total of 71,172 patients who underwent a spinal cord stimulator implant were included in the analysis, of which 52,070 underwent percutaneous and 19,102 underwent paddle spinal cord stimulator lead implant. The overall incidence of spinal cord injury after spinal cord stimulator implant (any lead type) was found to be 0.42% (302 patients). The incidence of spinal cord injury after percutaneous and paddle lead implants did not differ at 0.45% (233 patients) and 0.36% (69 patients)(p=0.12), respectively. Overall, variables associated with a significantly increased OR (95% confidence interval) of developing spinal cord injury included male gender by 1.31 times (1.04 to 1.65)(p=0.02); having a claim for low molecular weight heparin within 30 days by 3.99 times (1.47 to 10.82)(p<0.01); a diagnosis for osteoporosis within 1 year by 1.75 times (1.15 to 2.66)(p<0.01); and a diagnosis of cervical or thoracic spinal canal stenosis within 1 year by 1.99 (1.37 to 2.90)(p<0.001) and 4.00 (2.63 to 6.09)(p<0.0001) times, respectively.
CONCLUSIONS: Overall, our results support the notion that spinal cord stimulator implant continues to be a safe procedure for chronic pain patients. However, risk factor mitigation strategies for the prevention of spinal cord injury after spinal cord stimulator implant should be undertaken prior to performing the procedure.
METHODS: The PearlDiver-Mariner database of national all payer claims was used to identify patients who underwent spinal cord stimulator implant (percutaneous or paddle) and developed subsequent spinal cord injury within 45 days. The primary outcome was to determine the overall incidence of spinal cord injury after spinal cord stimulator implant. Secondary outcomes included an evaluation of potential factors associated with developing spinal cord injury using univariable and multivariable regression analysis.
RESULTS: A total of 71,172 patients who underwent a spinal cord stimulator implant were included in the analysis, of which 52,070 underwent percutaneous and 19,102 underwent paddle spinal cord stimulator lead implant. The overall incidence of spinal cord injury after spinal cord stimulator implant (any lead type) was found to be 0.42% (302 patients). The incidence of spinal cord injury after percutaneous and paddle lead implants did not differ at 0.45% (233 patients) and 0.36% (69 patients)(p=0.12), respectively. Overall, variables associated with a significantly increased OR (95% confidence interval) of developing spinal cord injury included male gender by 1.31 times (1.04 to 1.65)(p=0.02); having a claim for low molecular weight heparin within 30 days by 3.99 times (1.47 to 10.82)(p<0.01); a diagnosis for osteoporosis within 1 year by 1.75 times (1.15 to 2.66)(p<0.01); and a diagnosis of cervical or thoracic spinal canal stenosis within 1 year by 1.99 (1.37 to 2.90)(p<0.001) and 4.00 (2.63 to 6.09)(p<0.0001) times, respectively.
CONCLUSIONS: Overall, our results support the notion that spinal cord stimulator implant continues to be a safe procedure for chronic pain patients. However, risk factor mitigation strategies for the prevention of spinal cord injury after spinal cord stimulator implant should be undertaken prior to performing the procedure.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app