Characteristics of the ventilator pressure- and flow-trigger variables.
Intensive Care Medicine 1995 Februrary
Pressure- and flow-triggering are available in the Puritan Bennett 7200ae and Siemens SV 300. Using a mechanical lung model, we described the characteristics of the pressure- and flow-triggered continuous positive airway pressure (CPAP) of both ventilators. In the Puritan Bennett 7200ae, the pressure-triggered CPAP is characterized by the relatively insufficient flow delivery after the triggering, resulting in a greater lung pressure-time product (total PTP) than the flow-triggered CPAP. Pressure support of 5 cmH2O results in total PTP less than that with flow-triggered CPAP. In the Siemens SV 300, total PTP with pressure- or flow-triggered CPAP is comparable. Total PTP is less with pressure- or flow-triggered CPAP of the Siemens SV 300 than that of the Puritan Bennett 7200ae, respectively. The application of small pressure- or flow-triggered pressure support in the Puritan Bennett 7200ae eliminates the difference. The impact of these differences on patient inspiratory muscle work remains to be determined.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app