CLINICAL TRIAL
JOURNAL ARTICLE
RANDOMIZED CONTROLLED TRIAL
Add like
Add dislike
Add to saved papers

Radiographic evaluation of osseous displacement following intra-articular fractures of the distal radius: reliability of plain radiography versus computed tomography.

This study evaluated the reliability of plain radiography versus computed tomography (CT) for the measurement of small (< 5 mm) intra-articular displacements of distal radius fracture fragments. The plain radiographs and CT scans of 19 acute intra-articular distal radius fractures were used by 5 independent observers, using 2 standardized techniques, to quantify incongruity of the articular surface in a blinded and randomized fashion. Repeat measurements were performed by the same observers 2-4 weeks later, allowing determination of intraclass correlation coefficients (ICC) as a measure of intraobserver and interobserver agreement. The average maximum gap displacement on plain radiographs was 2.1 mm (range, 0.0-15.0 mm, lateral view) and on CT images was 4.9 mm (range, 0.7-17.3 mm, axial view). The average maximum step displacement on plain radiographs was 0.9 mm (range, 0.0-6.4 mm, lateral view) and on CT images was 1.2 mm (range, 0.0-6.0 mm, sagittal view). More reproducible values determining step and gap displacement were obtained when the arc method of measurement was used on CT scans (ICC values, .69-.97) as compared to the longitudinal axis method for plain radiographs (ICC values, .30-.50). For measured displacements of 2 mm or more, our data demonstrated poor correlation between measurements made on CT images and those made on plain radiographs (gap or step displacement > 2 mm, K = 0.21; step displacement > 2 mm, K = 0.21). Thirty percent of measurements from plain radiographs significantly underestimated or overestimated displacement compared to CT scan measurements. From these data, we conclude that CT scanning data, using the arc method of measurement, are more reliable for quantifying articular surface incongruities of the distal radius than are plain radiography measurements.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app